MBR-1404GMBR-1404GRBlood Bank RefrigeratorINSTRUCTION MANUAL MBR-1404GR
9 REFRIGERATOR COMPONENTS MBR-1404GR 1 2 3 4567891011 12 13 14 15 1617
10 REFRIGERATOR COMPONENTS 1. Front cover: Open this cover to connect to a remote alarm or when cleaning the condenser filter. Refer to page 1
11 REFRIGERATOR COMPONENTS Inside of front cover 1. Condenser filter: This filter is to prevent dust from plugging the c
12 REFRIGERATOR COMPONENTS Control panel 1. Alarm buzzer stop key (BUZZER): Press this key to temporarily silence the buzz
13 INSTALLATION Installation site To operate this unit properly and to obtain maximum performance, install the unit in a location with the fo
14 INSTALLATION Installation 1. Remove the packaging materials and tapes Remove all transportation packaging materials and tapes. Open the do
15 BEFORE COMMENCING OPERATION Preparation of monitor bottle Prior to the operation of the unit, fill the monitor bottles (upper and lower)
16 START-UP OF UNIT Follow the procedures for the initial and consequent operations of the unit. 1. Connect unit to dedicated power supply.
17 OPERATING INSTRUCTIONS Temperature display The unit is set at the factory to obtain the proper temperature automatically (4oC±1.5oC). The
18 OPERATING INSTRUCTIONS Alarm, safety function, and self diagnostic function This unit has the alarm and safety functions shown in Table
1 Note: 1. No part of this manual may be reproduced in any form without the expressed written permission of SANYO. 2. The contents of this manual
19 ROUTINE MAINTENANCE WARNING Always disconnect the power supply to the unit prior to any repair or maintenance of the unit in order to prevent
20 ROUTINE MAINTENANCE Cleaning of evaporating tray 1. Remove the evaporating tray installed on the back of the frame by lifting the tray.
21 ROUTINE MAINTENANCE Calibration of temperature sensor The temperature sensors are attached to the monitor bottles. Follow the followi
22 TROUBLESHOOTING If the unit malfunctions, check out the following before calling for service. If nothing operates even when switched on Th
23 SPECIFICATIONS Name Blood Bank Refrigerator Model MBR-1404G MBR-1404GR External dimensions W1440 x D830 x H1950 (mm) Internal dimensions
24 CAUTION Please fill in this form before servicing. Hand over this form to the service engineer to keep for his and your safety. Safety ch
7FB6P101236004 (5 April 2005) Recycled paper SANYO Electric Biomedical Co., Ltd.10-15, Hongo 3-Chome Bun
2 PRECAUTIONS FOR SAFE OPERATION It is imperative that the user complies with this manual as it contains important safety advice. Items and proc
3 PRECAUTIONS FOR SAFE OPERATION Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to rain
4 PRECAUTIONS FOR SAFE OPERATION If this unit is to be used for storing poisons, radioactive material or other harmful products, ensure that
5 PRECAUTIONS FOR SAFE OPERATION Select a level and sturdy floor for installation. This precaution will prevent the unit from tipping. I
6 PRECAUTIONS FOR SAFE OPERATION Dispose of any water in the evaporation tray completely before moving the unit. Spilled or splashed water m
7 CAUTIONS FOR USAGE 1. Adequate space should be provided between the items inside the unit to allow air circulation. 2. Never store corrosive m
8 ENVIRONMENTAL CONDITIONS This equipment is designed to be safe at least under the following conditions (based on the IEC 1010-1): 1. Indoor us
Komentarze do niniejszej Instrukcji