Podsumowanie treści

Strona 1 - MBR-1404GR

MBR-1404GMBR-1404GRBlood Bank RefrigeratorINSTRUCTION MANUAL MBR-1404GR

Strona 2 - CONTENTS

9 REFRIGERATOR COMPONENTS MBR-1404GR 1 2 3 4567891011 12 13 14 15 1617

Strona 3

10 REFRIGERATOR COMPONENTS 1. Front cover: Open this cover to connect to a remote alarm or when cleaning the condenser filter. Refer to page 1

Strona 4

11 REFRIGERATOR COMPONENTS Inside of front cover 1. Condenser filter: This filter is to prevent dust from plugging the c

Strona 5

12 REFRIGERATOR COMPONENTS Control panel 1. Alarm buzzer stop key (BUZZER): Press this key to temporarily silence the buzz

Strona 6

13 INSTALLATION Installation site To operate this unit properly and to obtain maximum performance, install the unit in a location with the fo

Strona 7

14 INSTALLATION Installation 1. Remove the packaging materials and tapes Remove all transportation packaging materials and tapes. Open the do

Strona 8 - CAUTIONS FOR USAGE

15 BEFORE COMMENCING OPERATION Preparation of monitor bottle Prior to the operation of the unit, fill the monitor bottles (upper and lower)

Strona 9 - ENVIRONMENTAL CONDITIONS

16 START-UP OF UNIT Follow the procedures for the initial and consequent operations of the unit. 1. Connect unit to dedicated power supply.

Strona 10 - REFRIGERATOR COMPONENTS

17 OPERATING INSTRUCTIONS Temperature display The unit is set at the factory to obtain the proper temperature automatically (4oC±1.5oC). The

Strona 11

18 OPERATING INSTRUCTIONS Alarm, safety function, and self diagnostic function This unit has the alarm and safety functions shown in Table

Strona 12

1 Note: 1. No part of this manual may be reproduced in any form without the expressed written permission of SANYO. 2. The contents of this manual

Strona 13

19 ROUTINE MAINTENANCE WARNING Always disconnect the power supply to the unit prior to any repair or maintenance of the unit in order to prevent

Strona 14 - INSTALLATION

20 ROUTINE MAINTENANCE Cleaning of evaporating tray 1. Remove the evaporating tray installed on the back of the frame by lifting the tray.

Strona 15

21 ROUTINE MAINTENANCE Calibration of temperature sensor The temperature sensors are attached to the monitor bottles. Follow the followi

Strona 16 - REMOTE ALARM TERMINAL

22 TROUBLESHOOTING If the unit malfunctions, check out the following before calling for service. If nothing operates even when switched on Th

Strona 17 - START-UP OF UNIT

23 SPECIFICATIONS Name Blood Bank Refrigerator Model MBR-1404G MBR-1404GR External dimensions W1440 x D830 x H1950 (mm) Internal dimensions

Strona 18 - OPERATING INSTRUCTIONS

24 CAUTION Please fill in this form before servicing. Hand over this form to the service engineer to keep for his and your safety. Safety ch

Strona 19

7FB6P101236004 (5 April 2005) Recycled paper SANYO Electric Biomedical Co., Ltd.10-15, Hongo 3-Chome Bun

Strona 20 - ROUTINE MAINTENANCE

2 PRECAUTIONS FOR SAFE OPERATION It is imperative that the user complies with this manual as it contains important safety advice. Items and proc

Strona 21

3 PRECAUTIONS FOR SAFE OPERATION Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to rain

Strona 22

4 PRECAUTIONS FOR SAFE OPERATION If this unit is to be used for storing poisons, radioactive material or other harmful products, ensure that

Strona 23 - DISPOSAL OF UNIT

5 PRECAUTIONS FOR SAFE OPERATION Select a level and sturdy floor for installation. This precaution will prevent the unit from tipping. I

Strona 24 - PERFORMANCE

6 PRECAUTIONS FOR SAFE OPERATION Dispose of any water in the evaporation tray completely before moving the unit. Spilled or splashed water m

Strona 25 - Safety check sheet

7 CAUTIONS FOR USAGE 1. Adequate space should be provided between the items inside the unit to allow air circulation. 2. Never store corrosive m

Strona 26 - 10-15, Hongo 3-Chome

8 ENVIRONMENTAL CONDITIONS This equipment is designed to be safe at least under the following conditions (based on the IEC 1010-1): 1. Indoor us